The CT-8002S-EIS combines precision charge-discharge testing with cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS) in a single portable instrument. CV scans and impedance measurements can be run between cycles without moving the cell to a separate potentiostat, saving time in aging, interfacial resistance and electrode reaction studies.
Key Features
- 2 independent channels, portable benchtop housing (~3 kg)
- -5 V to +5 V operation (discharge down to negative voltage)
- 6 current ranges (0.1 / 1 / 5 / 10 / 50 / 100 mA), minimum output current 0.2 µA
- Voltage & current accuracy ±0.01% F.S., stability 0.02% F.S.
- Current response time ≤20 µs, 1000 Hz recording (1 ms interval)
- Cyclic voltammetry (CV) scans can be inserted into charge-discharge programs; scan rate 0.01 mV/s – 1 V/s
- Electrochemical impedance spectroscopy (EIS): 10 mHz – 200 kHz
- CC, CV, CP, CC-CV charge/discharge; DCIR, pulse simulation and GITT tests
- 24-bit AD / 16-bit DA resolution; automatic switching between current ranges
- Bipolar linear circuit: current crosses zero smoothly with no charge/discharge switching delay (manufacturer data)
- Four-wire connection, dual closed-loop constant current / constant voltage control
- Offline testing and power-failure protection (manufacturer data)
- Compact and lightweight: can be used together with other instruments for in-situ measurements
- Simultaneous CV and EIS on both channels; impedance measurement runs alongside charge-discharge testing
Technical Specifications
| Model | CT-8002S-5V100mA-EIS |
| Channels | 2 |
| Voltage range | -5 V – +5 V |
| Current ranges | 0.1 mA / 1 mA / 5 mA / 10 mA / 50 mA / 100 mA |
| Minimum output current | 0.2 µA |
| CV cut-off current | 0.1 µA / 1 µA / 5 µA / 10 µA / 50 µA / 0.1 mA (by range) |
| Voltage & current accuracy | ±0.01% F.S. |
| Stability | Voltage 0.02% F.S., power 0.04% F.S. |
| Output power | 0.5 W / channel |
| Current response / conversion time | ≤20 µs |
| Recording frequency | 1000 Hz (minimum interval 1 ms) |
| Cycles / steps | 1–65535 cycles, ≤255 steps |
| CV scan rate | 0.01 mV/s – 1 V/s |
| EIS frequency range | 10 mHz – 200 kHz |
| Resolution | AD 24-bit / DA 16-bit |
| Connection | Four-wire |
| Data export | Excel, TXT, CSV, PDF, graph |
| Cell connection | Alligator clip or polymer fixture |
| Power supply | AC 110 / 220 V ±10%, 50/60 Hz, 50 W |
| Dimensions / weight | 127 × 226 × 48 mm / ~3 kg |
Applications
- Cell diagnostics and aging studies with EIS
- Interfacial (SEI) resistance analysis
- Cyclic voltammetry and electrode reaction studies
- In-situ measurements in battery materials research
- Coin cell and small-cell R&D testing
The price shown is for one test unit in standard configuration; channel count, cabinet/rack configuration, software and accessories are confirmed at quotation stage.
No questions have been asked for this product yet.