The 100 mA version of the mini all-in-one system combines a 25 L constant temperature chamber with a 32-channel high-precision coin cell tester with four current ranges. Measuring from the microampere level up to 100 mA, it delivers precise, temperature-controlled data for both low-current research cells and higher C-rate tests.
| Model | WHW-25L-15C-5V100mA (32 channels) |
| Nominal volume / channels | 25 L / 32 channels |
| Internal size | 360 × 300 × 220 mm |
| External size | 440 × 550 × 390 mm |
| Weight | approx. 45 kg |
| Tester | 2 × BT-4016Q-5V100mA (2 × 16 channels), 5 V, 100 mA |
| Current ranges | 0.2 µA–0.1 mA / 0.1–1 mA / 1–10 mA / 10–100 mA |
| Voltage & current accuracy | ±0.01% F.S. |
| Temperature range | 15–60 °C |
| Temperature fluctuation | ≤±0.5 °C (no load, stable) |
| Temperature deviation | ±2.0 °C (no load, stable) |
| Heating / cooling time | 25→60 °C ≤50 min / 25→15 °C ≤60 min |
| Cooling / heating | Semiconductor (thermoelectric) module, DC fan, PID control |
| Inner / outer wall | SUS304 stainless steel / coated cold-rolled steel, polyurethane insulation |
| Equipment | Hollow tempered glass door, LED lighting, 2 cell fixture connection plates, touch-key digital controller |
| Communication | Ethernet; control unit and network switch included, controlled with Neware BTS software |
The price shown is for one test unit in standard configuration; channel count, cabinet/rack configuration, software and accessories are confirmed at quotation stage.