The system combines a titanium gas diffusion flow cell with a high-pressure constant-flow pump; electrochemical measurements are made while electrolyte passes through the cell under high pressure.
The pump is set from 0.01 to 50 ml/min in 0.01 ml/min steps and works up to 30 MPa (20 MPa above 20 ml/min). Two options: anode and cathode sharing one channel, or separate channels for anode and cathode.
| Flow range | 0.01–50.00 ml/min |
| Maximum pressure | 30 MPa (0.10–20 ml/min), 20 MPa (20.01–50 ml/min) |
| Pressure pulsation | ≤ 0.4 MPa |
| Cell body | Titanium, 50 × 50 mm, 1 ml |
| Power | AC 220 V, 50 Hz, 200 W |
| Weight / size | 12 kg / 420 × 300 × 175 mm |
| Type options | Shared Channel (Anode and Cathode), Separate Channels (Anode and Cathode) |
Do not exceed the stated maximum pressure and take appropriate safety precautions when working under pressure.
Dimensions, channel sizes and volume can be customised on request; please state special requirements in your quotation request.