Select Language
Currency Selection
  • Jacketed Spherical Corrosion Test Cell
  • Jacketed Spherical Corrosion Test Cell

Jacketed Spherical Corrosion Test Cell

Product code : LM-EC-CORR-SPH-J         Comments (0)        Add Comment
Marka :  LabMateryal
Volume
500 mL 1000 mL
-
+

The spherical corrosion test cell is a round-body glass cell for measuring the corrosion behaviour of metal samples with three-electrode methods. A 15 mm diameter sample mounts on the working electrode holder; the exposed area is 1 cm² (manufacturer data).

An adjustable-angle salt bridge (Luggin capillary) points the reference electrode at the sample and reduces the ohmic (iR) drop. To keep chloride ions away from the sample, the salt bridge is filled with saturated potassium nitrate (KNO₃). The F-type inlet bubbles gas through the solution before the measurement and only feeds gas over the surface during it.

The set consists of the cell body, F-type gas inlet, salt bridge, liquid seal, PTFE stoppers, sample holder, reference electrode and graphite rod counter electrode (manufacturer data). The cell can be used in corrosive media other than HF at room and elevated temperature.

The body is made of borosilicate glass 3.3, which has low thermal expansion (3.3 × 10⁻⁶ K⁻¹) and the highest resistance classes against water (HGB 1) and acids (S 1). Avoid sudden temperature differences above 120 °C to reduce the risk of breakage.

In the jacketed (double-layer) version an outer water jacket surrounds the inner cell. Water from a circulating thermostatic bath flows through the jacket via hose connections and keeps the electrolyte at a constant temperature during the measurement.

Key Features

  • Jacketed (double-layer) body: constant electrolyte temperature with a circulating bath
  • Complete set with sample holder, reference and graphite counter electrode
  • Fixed 1 cm² sample area (Ø15 mm sample)
  • Adjustable-angle salt bridge with Luggin capillary
  • KNO₃-filled salt bridge for chloride-free measurements
  • 500 mL and 1000 mL options

Technical Specifications

Cell type Spherical corrosion cell, standard-sealed
Temperature control Water jacket for a circulating thermostatic bath
Body Borosilicate glass 3.3
Sample Ø15 mm, exposed area 1 cm²
Set contents Body, F-type inlet, salt bridge, liquid seal, PTFE stoppers, sample holder, reference electrode, graphite rod
Operating media Corrosive media except HF; room and elevated temperature
Volume options 500 mL, 1000 mL

For volumes, dimensions or port layouts other than those listed, you can request a custom-made quotation through the Quotation Form.

Applications

  • Potentiodynamic polarization and Tafel analysis
  • Corrosion rate by linear polarization resistance (LPR)
  • Electrochemical impedance spectroscopy (EIS)
  • Pitting and passivity studies
★
★
★
★
★
Total 0 Comment made


Thank you for your comment. Your comment will be published after it is done.
Comment on you
Verification Code


Thank you for your question. You will be notified by email once your question has been answered.
Click to Ask a Question
  No questions have been asked for this product yet.
E-Newsletter